Extraction of the AstraZeneca compound (NAE-111) using BioTrap 500 C18 / MS-detection

The BioTrap 500 C18 online-extraction column can be used with MS-detection, which is demonstrated in the chromatogram below.

Courtesy Astra Arcus, Sweden

Sample: NAE-111 (Astra compound)
Inj. vol.: 50 µl serum
Extraction column : BioTrap 500 C18, 20x4.0 mm
Extraction mobile phase: 2.5% 2-propanol in 10 mM sodium phosphate buffer, pH 7.0
Flow: 0.8 ml/min
Analytical column: Hypersil BDS C18, 100 x 4.6 mm, 3 µm
Analytical mobile phase: 17% acetonitrile, 0.1% formic acid
Flow: 1.0 ml/min
Detection: MS, APCI

Analysis Program

Switching valve position Analysis step Time
1. Extraction position sample extraction 3 min
2. Elution position transfer of analyte 3 min
3. Extraction position separation and reequilibration 3 min